Customization: | Available |
---|---|
Classification: | Vet Patient Monitor |
Type: | Vet Patient Monitor |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
2022 new Veterinary Patient monitor Vet monitor animal use monitor with ETCO2
Multi‐parameter patient monitor
Standard parameters: | 3/5‐Lead ECG, ECG, NIBP, SpO2, HR/PR, 2‐RESP, 2‐TEMP, Rechargeable Battery |
Optional configuration: | Built-in printer, SpO2, Nellcor SpO2, Masimo SpO2, 2-IBP, ETCO2(mainstream or side), WLAN accessory, Wall Mount, Moving cart |
ECG | |
Number of Leads : | 3 or 5 leads Lead View |
Lead output: | 3 wires/5 wires ECG cable |
User Slectable: | I,II,III,aVR,aVL,aVF,V(5 Lead);I,II or III(3 lead) |
Gain Slection: | x1/4,x1/2,x1 and x2 Frequency Response |
Diagnostic: | 0.05 to 90HZ |
Monitor: | 0.5 to 35 HZ |
Surgery: | Surgery:1-15HZ |
Calibration Signa : | 1 (mV p-p) |
Accuracy : | ECG calibration: ± 1 mv Heart rate accuracy: ± 1% |
ECG Signal Range: | ±8 m V ( Vp-p ) |
Review | Available |
NIBP (Non-invasive blood pressure) | |
Technique : | Oscillometric during inflation Range |
Measurement type: | Adult, pediatric, neonate |
Measurement parameter: | Systolic, Diastolic, Mean |
Adult: | 40-270mmHg |
Pediatric: | 40-240mmHg |
Neonate: | 10-150mmHg |
Measurment cycle: | <40 sec,typical |
Cycles(Selectable): | 2,3,5,10,15,30min 1,2,4hr |
STAT model: | 5 minutes of continuous reading Max Allowable Cuff Presure |
Adult: | 300 mmHg |
Pediatric: | 240mmHg |
Neonate: | 150 mmHg |
Resolutions: | 1mmHg |
Transducer Accurancy : | ±2mmHg or±2% of reading |
SpO2 | |
Display type: | Waveform, data |
Measurement range: | 0~100% |
Resolution: | 1 BPM |
Pulse rate range: | 30~250bpm |
Accuracy: | ±1bpm |
CO2(option) | |
Measurement range: | 0 ~ 10.0% ( Carbon dioxide inhalation and carbon dioxide exhalation) |
Accuracy: |
±2mmHg(<5% Measured values) 10%(>5% Measured values) |
Pulse Oxemetry | |
Range: | 0 to 100% |
Resolution: | 1% |
Accuracy: | 70% to 99% range ±2% ,1% to 69%range ±5% |
Bodymark: |
•Over 144 bodymark patterns for all application |
Method: | Dual wavelength LED |
Respiration Rate: | |
Rate: | Adult: 0 ~ 120 rpm Neo/Ped 0 ~ 150 rpm |
Resolution: | 1rpm |
Accuracy: | ± 2 breath/min |
Heart(Pulse) Rate | |
Range: | Adult 15-300 bpm,Pediatric 15-350 bpm |
Accurancy: | ±2bpm or ±3% whichever is greate |
Trends | |
Memory: | 96 hr at 30 second intervals for a NIBP reading every 5 mintues |
Display : | trabular;graphical |