Syinge Size 2 ml,5 ml, 10 ml, 20 ml, 30 ml, 50/60 ml
Volume Range 0.01 ~ 9999 ml, step by 0.01ml
Flow Rate Range
2ml: 0.01 ~ 100 mlh; 5ml: 0.01 ~ 150ml/h;
10 ml: 0.01~ 400 m/h; 20 ml: 0.01~ 600ml/h;
30 ml: 0.01~ 1000 m/h; 50/60ml: 0.01 ~ 2100m/h
Accuracy +-2%
Bolus Rate 2 ml Syringe: 100 mI/h;5 ml Syringe: 150 ml/h;
10 ml Syringe: 400 ml/h;20 ml Syringe: 600 ml/h;
30 ml Syringe: 1000 ml/h;50/60 ml Syringe: 2100 ml/h
KVO Rate
Flow rate z 10mlh: KVO rate 3 ml/h
Flow rate 21ml/h and <10m/h: KVO rate 1 ml/h
Flow rate < 1m/h: KVO rate = Flow rate
Alarm
Battery exhausted alarm,Low battery alarm,Near end alarm,VTBI finished alarm,
Incorrect syringe handle installation alarm,Syringe size error, Plunger error,
Pause overtime alarm,Occlusion alarm,Malfunction alarm,A.C. and D.C. Off
Occlusion Pressure
Maximum 130 kPa, Minimum 26 Kpa.
9 adjustable setting
Power Adapter A.C. Input power 100 V ~ 240 V 50/60 HZ; D.C. output power 15 V/ 2A
Lithium Battery Nominal Voltage: 10.8 V; Device can work over 6 hrs on battery
Infusion Pump Input Power D.C.15 V
Power <55VA
Classification Class I, Type CF, IPX2
Weight 1.5 kg
Dimension 270x 140 x 80 mm (WxD xH) |